• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GMBH MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GMBH MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM Back to Search Results
Model Number 03P6822
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/16/2023
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
 
Event Description
The customer observed a false elevated magnesium result generated on architect c4000 analyzer for one sample.Results provided: customer normal range 1.6 ¿ 2.6 mg/dl.(b)(6) 2023, sid (b)(6), initial result 7.8 mg/dl, repeat result on same analyzer 1.8 mg/dl no impact to patient management was reported.
 
Manufacturer Narrative
After further evaluation, the suspect medical device was changed from magnesium reagent kit, list number 03p68-22, abbott gmbh, max-planck-ring 2, wiesbaden, germany, 65205 to architect c4000, list number 02p24-40, abbott laboratories, 1915 hurd drive, irving, tx, 75038.Mdr number 3016438761-2023-00097 has been submitted and all further information will be documented under that mdr number.H3 other text : after further evaluation, the suspect medical device was changed from magnesium reagent kit, list number 03p68-22 to architect c4000, list number 02p24-40.Mdr number 3016438761-2023-00097 has been submitted and all further information will be documented under that mdr number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGNESIUM
Type of Device
PHOTOMETRIC METHOD, MAGNESIUM
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key16302835
MDR Text Key308863175
Report Number3002809144-2023-00072
Device Sequence Number1
Product Code JGJ
UDI-Device Identifier00380740161620
UDI-Public00380740161620
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181748
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number03P6822
Device Catalogue Number03P68-22
Device Lot Number46893UN22
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/23/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ARC C4000 INTGR, 02P24-40, C402081; ARC C4000 INTGR, 02P24-40, C402081
-
-