• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS NORTH AMERICA LLC HEARTSTART XL+ DEFIBRILLATOR/MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS NORTH AMERICA LLC HEARTSTART XL+ DEFIBRILLATOR/MONITOR Back to Search Results
Model Number 861290
Device Problem Charging Problem (2892)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/10/2023
Event Type  malfunction  
Event Description
Philips received a complaint on the heartstart xl+ defibrillator indicating that the defibrillation test failed.There was reportedly no patient involvement.The rse worked with site in evaluating the device.It was determined that this was a malfunction of the high voltage capacitor, which was sent to the customer for installation.The device remains at the customer site and no further evaluation is warranted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART XL+ DEFIBRILLATOR/MONITOR
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
daniel derochers
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key16303720
MDR Text Key308850315
Report Number3030677-2023-00529
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00884838023680
UDI-Public00884838023680
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K110825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number861290
Device Catalogue Number861290
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/10/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-