It was reported that the patient presented in clinic for follow-up after a syncopal event.Upon review, the implantable cardioverter defibrillator was found on backup mode.The device could not be restored as communication with the programmer was lost.The device was not pacing in backup mode, and the patient was uncomfortable with bradycardia as a result.The device was explanted and replaced.The patient was recovering.
|
The reported field event of backup vvi, telemetry anomaly, and no pacing output was confirmed.Analysis of device image found the device went into backup mode due to a power-on-reset (por).The por, telemetry anomaly, and no pacing output were caused by low battery voltage.The low battery voltage was due to device usage resulting from the patient's repetitive high voltage therapies delivered in a short period of time.After the device was restored from backup mode, functional testing was performed and found normal.
|