• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESMED LTD ASTRAL 150 - AMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESMED LTD ASTRAL 150 - AMER Back to Search Results
Model Number 27003
Device Problems Defective Device (2588); Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2022
Event Type  malfunction  
Manufacturer Narrative
The astral device was returned to a resmed authorized third party service center for evaluation and service.No further information is available at this time.Based on all available evidence, an investigation determined that the reported complaint was due to faulty/defective internal battery.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference#: (b)(4).
 
Event Description
It was reported to resmed that an astral device displayed an error message (sf180) related to a battery charger fault.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 - AMER
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
MDR Report Key16326178
MDR Text Key309131109
Report Number3007573469-2023-00093
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00619498270033
UDI-Public(01)00619498270033(11)200823(10)1476889
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation 505
Type of Report Initial
Report Date 02/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27003
Device Catalogue Number27003
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA02/08/2023
Distributor Facility Aware Date01/10/2023
Device Age28 MO
Date Report to Manufacturer02/08/2023
Date Manufacturer Received01/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-