It was reported there was csf leakage with an evd transducer.Patient had evd placed on (b)(6) 2023.The leakage was noticed when patient was having an emergent scan due to significant neuro decline.Edwards caps were replaced with hospital owned caps to resolve the issue.No additional troubleshooting was completed as the patient was going to emergent ct and no additional treatment needed for leakage.Nurse is unsure if neurosurgery replaced the entire system, but commented that it is protocol to replace evd system when sterility is broken.Per follow up with sales rep, he has visited site multiple times and each time rep cannot find anything in their practice that would cause leakage.Lot number is unknown and device is unable to be sent back.
|