The initial reporter stated they had quality control issues for multiple tests and received results above or below the measuring range for an unspecified number of patient samples tested with multiple elecsys assays on a cobas e 801 analytical unit.The field service engineer checked and repaired the analyzer and the samples were repeated on (b)(6) 2023.The customer provided an example of two patient samples, tested with the elecsys tsh and elecsys ft4 assays.Patient sample 1, tested with elecsys tsh: the sample initially resulted in a tsh value of < 0.007 uiu/ml, accompanied by a data flag.On (b)(6) 2023, the sample was repeated after the repair and resulted in a tsh value "between 2.0 - 4.0 uiu/ml".No specific value was provided.Patient sample 2, tested with elecsys ft4: the sample initially resulted in an ft4 value of > 7.77 ng/dl, accompanied by a data flag.On (b)(6) 2023, the sample was repeated after the repair and resulted in an ft4 value of 3.9 ng/dl.The customer further mentioned that elecsys ft3, vitamin d, vitamin b12, and folate also recovered above the measuring range initially and "repeated anywhere from 25-50% lower" on repeat after the repair.No specific values were provided."some" of the initial questionable results were reported outside of the laboratory and an amended report was issued.The tsh, ft4, ft3, vitamin d, vitamin b12 and folate reagent lots and expiration dates were requested but not provided.
|