ST. JUDE MEDICAL, INC. ENSITE VELOCITY¿ DWS7 DISPLAY WORKSTATION; COMPUTER, DIAGNOSTIC, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 100129557 |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/05/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results of the investigation were inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information received, the cause of the reported incident could not be determined.
|
|
Event Description
|
During an af procedure, it was noted that there were connectiong issues with the ensite dws resulting in multiple restarts which caused delays in treatment.It was noted that the ensite dws was unable to boot up and there was a flashing red light.The dws would close without prompt about 15 seconds after the start button was pressed.The dws was restarted at least five times, but the issue persisted.Cabling was checked, and it was discovered that if the hdmi cable was unplugged, the dws was able to boot.The procedure was completed without adverse patient consequences.
|
|