AESCULAP AG TSPACE XP IMPLANT 5° 30X11.5X8MM; TSPACE XP IMPLANTS Back to Search Results
Model Number SO938P
Device Problem Unintended System Motion (1430)
Patient Problem Laceration(s) (1946)
Event Date 01/10/2023
Event Type  Injury  
Event Description
It was reported to aesculap inc.That a tspace xp implant 5° 30x11.5x8mm ( product # so938p) was implanted during a transforaminal lumbar interbody fusion (tlif) procedure on (b)(6) 2023.According to the complainant, upon insertion of the trial implant using the articulating inserter; the trial defaulted to 90 degrees prior to entering the disc space, and bumped/tore the dura.Further information received stated that the torn dura was sutured closed.It was reported that the surgery was delayed approximately twenty (20) minutes.The implant was reported to not be available for evaluation.The adverse event/malfunction is filed under aic reference (b)(4).Associated reports: 400585236_2916714-2023-00028_mdr, 400588773_2916714-2023-00029_mdr.
 
Manufacturer Narrative
Investigation on-going.Should relevant additional information / investigation results become available, a supplemental report will be submitted.
 
Event Description
Investigation complete.
 
Manufacturer Narrative
Investigation results: the device was not returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation of the unit was performed and no parts were returned for failure analysis.Therefore, the investigation was not able to confirm a device issue that could be associated with the reported event.Batch history review: the device history records (dhr) were not able to be reviewed as no lot number was made available.However, all device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.Conclusion and measures / preventative measures: a definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TSPACE XP IMPLANT 5° 30X11.5X8MM
Type of Device
TSPACE XP IMPLANTS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key16340334
MDR Text Key309260075
Report Number2916714-2023-00027
Device Sequence Number1
Product Code MAX
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/09/2023,08/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/09/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSO938P
Device Catalogue NumberSO938P
Device Lot Number52340409
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA02/09/2023
Distributor Facility Aware Date01/11/2023
Device Age68 MO
Event Location Hospital
Date Report to Manufacturer01/11/2023
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age57 YR
Patient SexFemale
-
-