Model Number AZURION 7 M20
Device Problem Difficult or Delayed Activation (2577)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/31/2023
Event Type  malfunction  
Event Description
It has been reported to philips that the touch screen module and the control module turned off.No harm has been reported to philips.
 
Manufacturer Narrative
Philips has investigated this complaint.According to the additional information collected, the system was in clinical use when the issue occurred.The vascular surgery was completed using another system.The field service engineer (fse) inspected the system onsite and confirmed the reported problem.Upon functional testing fse identified power distribution unit (pdu) dc module had no power and determined that fault was with (pdu) dc module.The fse replaced pdu dc module.After this, the system was returned to use in good working order.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZURION
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key16342881
MDR Text Key309276685
Report Number3003768277-2023-00792
Device Sequence Number1
Product Code OWB
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K163715
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/09/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAZURION 7 M20
Device Catalogue Number722079
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/17/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-