• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORNIER S.A.S. AEQUALIS REVERSED GLENOSPHERE ECCENTRIC 36 MM X 25MM; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER S.A.S. AEQUALIS REVERSED GLENOSPHERE ECCENTRIC 36 MM X 25MM; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Model Number DWD182
Device Problems Use of Device Problem (1670); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/17/2023
Event Type  malfunction  
Manufacturer Narrative
Analysis results are not available at the time of this report.A follow-up report will be sent when the analysis is complete.
 
Event Description
It was reported that there was a glenosphere defect.Another implant was used to treat the patient.No impact on the patient.
 
Manufacturer Narrative
The reported event could be confirmed (jammed safety screw in the glenoid sphere after opening the package) since the device was returned and was found to be in condition stated in the event description.The device inspection revealed the following: the visual inspection confirmed that the safety screw was jammed in the glenoid sphere.After loosening the safety screw, the functionality of the sphere was checked, and it could correctly impact and tighten into the baseplate-gauge without issue to report.The issue described in the reported event was unable to be reproduced.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any additional information is provided, the investigation will be reassessed.
 
Event Description
It was reported that there was a glenosphere defect.Another implant was used to treat the patient.No impact on the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AEQUALIS REVERSED GLENOSPHERE ECCENTRIC 36 MM X 25MM
Type of Device
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR  38330
Manufacturer (Section G)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR   38330
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16343726
MDR Text Key309272469
Report Number3000931034-2023-00055
Device Sequence Number1
Product Code KWS
UDI-Device Identifier03700386935848
UDI-Public03700386935848
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K081059
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 04/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/09/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDWD182
Device Catalogue NumberDWD182
Device Lot Number9448AX
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-