• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT SBP DR INDUCTIVE PACEMAKER, JPD; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT SBP DR INDUCTIVE PACEMAKER, JPD; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM2152
Device Problems Premature Discharge of Battery (1057); Failure to Interrogate (1332); No Pacing (3268)
Patient Problems Fatigue (1849); Dizziness (2194)
Event Date 01/05/2023
Event Type  Injury  
Event Description
During an in clinic follow-up, it was not possible to interrogate the device.An electrocardiogram (ecg) was performed and the device was no longer pacing the patient.Premature depletion was suspected.The patient reported experiencing fatigue and dizziness.The device was explanted and replaced to resolve the event.The patient was stable.
 
Manufacturer Narrative
The reported events of inability to interrogate, no output and premature battery depletion were not confirmed.The device was received with normal telemetry and output.Analysis performed did not identify any functional issues.Visual inspection of the header attachment area detected an anomaly between the pre-molded header and titanium case.The device was cut open to enable further testing and the battery was found depleted.Feedthrough leak test was performed, indicating a device hermeticity breach.This is consistent with feedthrough damage as a result of fluid intrusion between the header and case, and subsequent fluid ingress to internal electronics.Hybrid circuitry was tested, and the results indicated normal current.A manufacturing process anomaly may have occurred, which resulted in the header bonding anomaly.The device is included in the zenex, assurity, endurity laser adhesion preparation advisory issued by abbott on 20 july 2022 for a subset of devices distributed and implanted outside of the united states.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SBP DR INDUCTIVE PACEMAKER, JPD
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16344148
MDR Text Key309254114
Report Number2017865-2023-05549
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/09/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2021
Device Model NumberPM2152
Device Lot NumberP000098086
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/04/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-