Model Number HL-90 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/20/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Udi section of device identification is unknown, no product information has been provided to date.Date of event is unknown.No information has been provided to date.Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that the fluid warmer display was jumping to values way out of range.There has been no report of patient involvement, no observable clinical symptoms, or a change in symptoms identified in the patient.
|
|
Event Description
|
Additional information received informing us that the issue was discovered during routine maintenance.There were no patients or providers harmed.
|
|
Manufacturer Narrative
|
One warmer device was received for investigation.During visual inspection the device was observed to have a corroded drain fitting, cracked tank cover, bent micro switch, and damaged display.The reported issue was confirmed during functional testing when the device temperature display was observed to fluctuate.The issue was traced to the device thermistor.Which was determined to be faulty.However, the investigation could not attribute a root cause for the observed condition of the thermistor.As no manufacturing root cause could be identified, no review of manufacturing device history records was conducted.A review of device confirmed this device has not been in for service in the previous year and there was no indication of a service issue during the investigation.Due to the age of the device, it has been deemed beyond economical repair and will be decommissioned.
|
|