Model Number HL-90
Device Problem Device Displays Incorrect Message (2591)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/20/2023
Event Type  malfunction  
Manufacturer Narrative
Udi section of device identification is unknown, no product information has been provided to date.Date of event is unknown.No information has been provided to date.Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the fluid warmer display was jumping to values way out of range.There has been no report of patient involvement, no observable clinical symptoms, or a change in symptoms identified in the patient.
 
Event Description
Additional information received informing us that the issue was discovered during routine maintenance.There were no patients or providers harmed.
 
Manufacturer Narrative
One warmer device was received for investigation.During visual inspection the device was observed to have a corroded drain fitting, cracked tank cover, bent micro switch, and damaged display.The reported issue was confirmed during functional testing when the device temperature display was observed to fluctuate.The issue was traced to the device thermistor.Which was determined to be faulty.However, the investigation could not attribute a root cause for the observed condition of the thermistor.As no manufacturing root cause could be identified, no review of manufacturing device history records was conducted.A review of device confirmed this device has not been in for service in the previous year and there was no indication of a service issue during the investigation.Due to the age of the device, it has been deemed beyond economical repair and will be decommissioned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE BLOOD AND FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key16357523
MDR Text Key309379878
Report Number3012307300-2023-01160
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 04/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? No
Date Returned to Manufacturer01/25/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/29/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-