Model Number MMT-723RNAS
Device Problems Excess Flow or Over-Infusion (1311); Environmental Compatibility Problem (2929)
Patient Problem Hyperglycemia (1905)
Event Date 06/12/2021
Event Type  malfunction  
Event Description
Medtronic received information that exposed to magnetic field or mri, possible over delivery anomaly occurred.There was an allegation of hyperglycemia as a result of this event.
 
Manufacturer Narrative
This report is part of a retrospective review and remediation.(b)(4).Sv 3.0b unit received with all operating currents within spec and passed functional testing including the rewind, a21 error test, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and displacement test.No possible over delivery anomaly noted.Unit was downloaded to carelink pro properly and all bolus delivered were found at the carelink report.Pump received with cracked belt clip slot, cracked battery tube threads, stained end cap sticker, stained address/serial number label and cracked reservoir tube lip.Pump passed the dat test at 0.0875 inches.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAS
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*  00777-3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*   00777-3869
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
*   00777-3869
7635140379
MDR Report Key16358187
MDR Text Key309578315
Report Number2032227-2023-156415
Device Sequence Number1
Product Code OYC
UDI-Device Identifier000000643169513143
UDI-Public(01)000000643169513143
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 02/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-723RNAS
Device Catalogue NumberMMT-723RNAS
Device Lot NumberA1723RNASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/24/2021
Date Manufacturer Received06/15/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/23/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age71 YR
Patient Weight184
-
-