• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OSTEOMED, LLC LEVER ARM MOTOR UNIT; HANDPIECE, ROTARY BONE CUTTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OSTEOMED, LLC LEVER ARM MOTOR UNIT; HANDPIECE, ROTARY BONE CUTTING Back to Search Results
Model Number 450-0034
Device Problem Power Problem (3010)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
The device had been returned under a service work order (sro) during which it was identified that the device was continuously running due to malfunction of the flex circuit in the cable assembly.The device was repaired under the sro and was subsequently tested and met specifications.The device is a reusable device.Attempts are being made to obtain further information regarding the findings of the sro, and a follow up report will be submitted if additional information is received and/or after completion of the investigation.Mdr submission timeline affected by data entry timeliness of the device diagnosis within the diagnosis/repair log.This diagnosis/repair log is reviewed on a weekly basis, and the data entry oversight resulted in the diagnosis of this device not being observed on the log until the weekly review performed on 10 february 2023.Actions are being taken to mitigate the data entry timeliness.
 
Event Description
A service request for the device was received with no product complaint made.On 09 january 2023, the service request work order device diagnosis was completed.During the service request work order, it was noted the device was running continuously due to a malfunctioning flex circuit in the cable assembly.As a result of these findings of the service request work order, this report is being submitted.Additionally, as a result of the service request work order, the customer is being contacted to obtain additional information regarding these findings.
 
Manufacturer Narrative
Manufacturing and inspection records were reviewed.All inspections passed and no issues were identified in the device history record review.The device was returned per service work order (sro) s000053005 and subsequently tested after repair.All specifications were met.The diagnosis during the work order was identified to be "unit runs continuously due to a malfunctioning flex circuit in the cable assembly." the device is a reusable instrument and was manufactured in 2020; therefore, it is unknown how many times the device was used prior to this reported complaint event nor under what circumstances it was used.No additional information regarding this event has been received.Therefore, the root cause of the issue could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVER ARM MOTOR UNIT
Type of Device
HANDPIECE, ROTARY BONE CUTTING
Manufacturer (Section D)
OSTEOMED, LLC
3885 arapaho rd
addison TX 75001
Manufacturer (Section G)
OSTEOMED, LLC
3885 arapaho rd
addison TX 75001
Manufacturer Contact
ellie wood
3885 arapaho rd
addison, TX 75001
9726774600
MDR Report Key16359149
MDR Text Key309494880
Report Number2027754-2023-00003
Device Sequence Number1
Product Code KMW
UDI-Device Identifier00845694015646
UDI-Public(01)00845694015646(30)1(11)200916
Combination Product (y/n)N
Reporter Country CodeID
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number450-0034
Device Catalogue Number450-0034
Device Lot Number1150854
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2022
Date Manufacturer Received03/06/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-