Model Number CI24R (CS)
Device Problem Electrical Shorting (2926)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 01/19/2023
Event Type  malfunction  
Manufacturer Narrative
This report is submitted on february 13, 2023.
 
Event Description
Per the clinic, the patient experienced performance decrement with the device.Reprogramming attempts were made; however, the issue could not be resolved.The device was explanted (b)(6) 2023, and the patient was reimplanted with another cochlear device during the same surgery.
 
Manufacturer Narrative
Device analysis indicated device failure.Device analysis report is attached.This report is submitted on april 24, 2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUCLEUS 24 CONTOUR
Type of Device
NUCLEUS 24 COCHLEAR IMPLANT SYSTEM
Manufacturer (Section D)
COCHLEAR LIMITED
1 university avenue
macquarie university
nsw 2109
AS  2109
Manufacturer (Section G)
COCHLEAR LTD
1 university avenue
macquarie university, nsw 2109
AS   2109
Manufacturer Contact
yaneesa vetsandonphong
1 university avenue
no. 8 jalan kerinchi
macquarie university, nsw 2109
AS   2109
MDR Report Key16360186
MDR Text Key309367752
Report Number6000034-2023-00411
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P970051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberCI24R (CS)
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received04/18/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age23 YR
Patient SexMale
-
-