Catalog Number 705.52P
Device Problem Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2023
Event Type  malfunction  
Event Description
A physician reported that an ophthalmic scissors would not open or close during the surgery.The procedure details were not provided.The surgery was completed after replacing the product with another one.There was no patient harm.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
A review of the device history record (dhr) traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The sample is received with inner blister, outer blister and cover foil showing the lot information.The sample shows no macroscopic signs of damage.The sample shows signs of surgery residues and is returned in a opened status.The sample was visually inspected with the aid of a photomicroscope with various magnifications.The sample was functionally tested.It was noticed that the scissors get stuck in a closed status.The scissors only can get open again by pushing the shaft manually down.As the blister was opened by the customer, he handled the product.The point in time when the malfunction occurred, can no longer be determent.The customers complaint is confirmed.It cannot be determined how or where the damage to the sample occurred; therefore a root cause for the customers reported event could not be determined.However, the damage was consistent with damage that can occur due to improper handling.The exact root cause for the customers reported event is unknown, therefore, specific action cannot be taken.Complaints are reviewed and monitored at regular intervals for adverse trends.No adverse trends have been observed associated with the reported product and event.No action has been identified for this reported event.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRIESHABER REVOLUTION DSP SCISSORS
Type of Device
SCISSORS, OPHTHALMIC
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
Manufacturer (Section G)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ   8203
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16377286
MDR Text Key309561433
Report Number3003398873-2023-00024
Device Sequence Number1
Product Code HNF
UDI-Device Identifier07612717070691
UDI-Public07612717070691
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2023
Device Catalogue Number705.52P
Device Lot NumberF186698
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/31/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-