• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number HL 20
Device Problem Failure to Pump (1502)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/30/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation is ongoing.As soon as new information becomes available, a follow up medwatch will be submitted.
 
Event Description
It was reported that the hl20 device displayed the error message: "runaway" and the led was blinking during checking of the machine.No harm to any person was reported.Complaint id#(b)(4).
 
Event Description
Complaint id #(b)(4).
 
Manufacturer Narrative
It was reported that the hl20 pump displayed the error message: "runaway" and the led "l" was blinking during setup of the device.This is the battery charge indicator with "l" on the left side of the console under the main switch and it is blinking when there is a problem with the battery.A getinge field service technician (fst) was onsite for investigation of the pump in question.The fst adjusted the settings of the optical tacho voltage and replaced the batteries.After that the pump was working to factory specification.Device was manufactured in 2017-02-07 (mat#701043261).The review of the non-conformities during the period of 2017-02-07 to 2023-01-30 does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences are unlikely.According to the service manual heart-lung machine hl 20 version 02 in chapter 6.11.6 adjustment of tpm optical tacho board the following is stated: the optical tacho must be adjusted after replacement or in case of speed differences between control system and safety system.The most probable root cause for the reported error message runaway could be determined as aging.Aging can change the properties of electronic components, which can cause it to have to be re-adjusted.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key16384915
MDR Text Key309642941
Report Number8010762-2023-00087
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL 20
Device Catalogue Number701028580
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/07/2017
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-