BOSTON SCIENTIFIC CORPORATION EASYTRAK 2 IS-1; IMPLANTABLE LEAD Back to Search Results
Model Number 4542
Device Problems Fracture (1260); High impedance (1291); High Capture Threshold (3266)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/02/2023
Event Type  Injury  
Event Description
It was reported that this left ventricular (lv) lead was explanted due to a fracture.It was noted that prior to procedure there were high impedance and threshold measurements; subsequently, multi-vector testing was performed without success.A lv lead from a different manufacturer was used as replacement.No additional adverse patient effects were reported.As of today, this product has not been received by boston scientific for further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2 IS-1
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16386829
MDR Text Key309667342
Report Number2124215-2023-07124
Device Sequence Number1
Product Code NIK
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P010012/S024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/16/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/24/2014
Device Model Number4542
Device Catalogue Number4542
Device Lot Number155924
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/02/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/24/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age45 YR
Patient SexFemale
-
-