Model Number EMAX2PLUS |
Device Problems
Vibration (1674); Environmental Particulates (2930); Noise, Audible (3273); Excessive Heating (4030)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the reported condition of the device smoking was not confirmed.Therefore, an assignable root cause was not determined.However, the device generating heat and vibrating identified during service and evaluation were confirmed.The assignable root cause was determined to be due to component failure from normal wear.Udi: (b)(4).
|
|
Event Description
|
It was reported by japan that during service and evaluation, it was determined that the motor device had vibration and was generating heat.It was further observed that the device had component, flex circuit and cord damage, insufficient/low power, and was making an excessive noise.It was determined that the labeling had an illegible etch.It was further noted that the hose was scratched, the cap wire was broken, the electric had leakage and the device had an insufficient rotational speed.It was further determined that the device failed pretest for visual assessments, cable assessment, motor thermistor assessment, no short circuit between phases and connector body, no short circuit between hall sensors and connector body, rotational speed assessment, rotational speed assessment, noise assessment and handpiece temperature assessment.It was noted in the service order that the device made an abnormal noise and was smoking.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in (b)(6) 2023.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|