Model Number 71992-01 |
Device Problems
Unable to Obtain Readings (1516); Device Displays Incorrect Message (2591)
|
Patient Problems
Hypoglycemia (1912); Chills (2191); Loss of consciousness (2418)
|
Event Date 02/03/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device mfg date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
An error message was reported with the adc device.Customer received a "replace sensor" message and was unable to obtain readings.As a result, customer experienced feeling bad, cold and loss of consciousness.Customer self-treated with sugar and denied third-party medical intervention.There was no report of death or permanent impairment associated with this event.
|
|
Manufacturer Narrative
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.No product has been returned and a valid serial number has not been provided.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.Clinical data was reviewed and confirmed that libre sensors continue to be safe, effective, and perform as intended in the field.A tripped trend review was conducted for the reported complaint and fs libre sensors, no trends were identified that would indicate any product related issues.If the product is returned, a physical investigation will be performed, and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
An error message was reported with the adc device.Customer received a "replace sensor" message and was unable to obtain readings.As a result, customer experienced feeling bad, cold and loss of consciousness.Customer self-treated with sugar and denied third-party medical intervention.There was no report of death or permanent impairment associated with this event.
|
|
Search Alerts/Recalls
|