• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LENSX, INC. LENSX LASER SYSTEM; OPHTHALMIC FEMTOSECOND LASER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LENSX, INC. LENSX LASER SYSTEM; OPHTHALMIC FEMTOSECOND LASER Back to Search Results
Model Number 550
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Capsular Bag Tear (2639)
Event Date 01/25/2023
Event Type  Injury  
Event Description
A customer reported experienced a posterior capsule tear and perform a vitrectomy in right eye of a patient, during surgery.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.Alcon lensx (site # (b)(4) is no longer operational.Lensx manufactured products are maintained and investigated by the alcon research, ltd.Irvine technology center site # (b)(4).The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
Additional information provided in h.3., h.6., and h.10.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Service history was reviewed for the system.No service record relevant to the complaint reported event was found.However, the system was last serviced prior to the reported event per service record (sr).The system found to meet all cosmetic and performance standards.Based on the assessment, the product met release criteria.The customer reported event cannot be confirmed.Thus, based on the information obtained, the root cause of the reported event is inconclusive.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LENSX LASER SYSTEM
Type of Device
OPHTHALMIC FEMTOSECOND LASER
Manufacturer (Section D)
ALCON LENSX, INC.
33 journey
suite #175
aliso viejo CA 92658
Manufacturer (Section G)
ALCON LENSX, INC.
33 journey
suite #175
aliso viejo CA 92658
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16397458
MDR Text Key309762352
Report Number2028159-2023-00216
Device Sequence Number1
Product Code OOE
UDI-Device Identifier00380659981623
UDI-Public00380659981623
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163551
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/17/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/30/2023
Device Model Number550
Device Catalogue Number8065998162
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/30/2013
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexMale
-
-