• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ACON LABORATORIES, INC. FLOWFLEX COVID-19 ANTIGEN HOME TEST; ANTIGEN DIAGNOSTIC FOR SARS-COV-2, PRODUCT CODE: QKP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ACON LABORATORIES, INC. FLOWFLEX COVID-19 ANTIGEN HOME TEST; ANTIGEN DIAGNOSTIC FOR SARS-COV-2, PRODUCT CODE: QKP Back to Search Results
Lot Number COV2020166
Device Problem No Device Output (1435)
Patient Problem Insufficient Information (4580)
Event Date 01/03/2023
Event Type  malfunction  
Event Description
Invalid result (s).No result developed.The user performed the test procedure correctly.
 
Manufacturer Narrative
Batch records, final product manufactured and qc records were reviewed for lot cov2020166.No abnormal issue was found in the manufacturing process, technical testing and quality control inspection, the manufacturing process complied with the dmr.Test results of retention saamples met with the qc criteria, the reported issue was not found or could not be reproduced.This complaint is not verified.In this follow-up report, the following information has been updated from the initial report upon completion of the internal investigation.The following fields are updated: b4, "date of this report" - date of follow-up report.G6, "type of report" - follow-up #1.H2, "if follow-up, what type?" - updated to "additional information" and "device evaluation" h6 "adverse event problem" - event problem and evaluation codes such as "investigation findings" and "investigation.Conclusions" are added per investigation.H10 "additional narrative/data" - added manufacturer's narrative.
 
Event Description
Invalid result (s).No result developed.The user performed the test procedure correctly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOWFLEX COVID-19 ANTIGEN HOME TEST
Type of Device
ANTIGEN DIAGNOSTIC FOR SARS-COV-2, PRODUCT CODE: QKP
Manufacturer (Section D)
ACON LABORATORIES, INC.
5850 oberlin drive #340
san diego CA 92121
Manufacturer (Section G)
ACON LABORATORIES, INC.
5850 oberlin drive #340
san diego CA 92121
Manufacturer Contact
qiyi xie
5850 oberlin drive #340
san diego, CA 92121
8588758011
MDR Report Key16397978
MDR Text Key309773698
Report Number2531491-2023-00482
Device Sequence Number1
Product Code QKP
Combination Product (y/n)N
PMA/PMN Number
EUA210494
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/17/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/23/2023
Device Lot NumberCOV2020166
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received01/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-