• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH LRG TAP PRI MOD NCK 16DEG 34MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS, CALICU

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH LRG TAP PRI MOD NCK 16DEG 34MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS, CALICU Back to Search Results
Catalog Number NLS-341600P
Device Problems Corroded (1131); Device-Device Incompatibility (2919)
Patient Problem Metal Related Pathology (4530)
Event Date 06/01/2014
Event Type  Injury  
Event Description
It was reported that the patient had cobalt level 1.2 states high - hip has popped out of socket at times.High level of cobalt, dislocation, corrosion, revision.It was reported that the patient's right hip was revised due to recurrent dislocations.Intra-operatively, corrosion was noted at both the stem-neck interface and head-trunnion interface.All components were removed.Rep provided pictures and confirmed that no further information will be released by the hospital or surgeon.
 
Manufacturer Narrative
An event regarding abnormal ion level and corrosion involving a rejuvenate modular device was reported.The event was confirmed.Method & results: device evaluation and results: visual inspection of images note blood drops and black material on neck surface.Device history review: review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: the complaint databases show there have not been other events for the reported lot.Similar events have occurred for the catalog number and product family.These events were determined to be associated with ra 2012-067.Conclusions: voluntary recall ra 2012-067 was initiated for abgii and rejuvenate modular stems and necks due to the potential risks associated with these devices.The reported high level of cobalt is considered to be under the scope of this recall.No further investigation is required.Not returned to the manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LRG TAP PRI MOD NCK 16DEG 34MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS, CALICU
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16406767
MDR Text Key309881090
Report Number0002249697-2023-00175
Device Sequence Number1
Product Code MEH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092561
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date 02/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/21/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/31/2015
Device Catalogue NumberNLS-341600P
Device Lot Number31715101
Was Device Available for Evaluation? No
Date Manufacturer Received01/26/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/07/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-2090-2012
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age73 YR
Patient SexMale
-
-