MAUDE Adverse Event Report: PULSION MEDICAL SYSTEMS SE PICCO CATHETER; PROBE, THERMODILUTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PULSION MEDICAL SYSTEMS SE PICCO CATHETER; PROBE, THERMODILUTION Back to Search Results
Model Number PV2014L22-A
Device Problem Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/27/2022
Event Type  Injury  
Manufacturer Narrative
It has been reported that as the guidewire was withdrawn through the cannula, the guidewire was split.No problems have been encountered during insertion of the guidewire.But the guidewire could only be withdrawn against resistance.A new catheter has been placed with a new, non defective guidewire.Only the guidewire has been returned.Inspection of the returned product has been performed.The microscopic inspection revealed that the guidewire has not only been split on one end, but it has also been torn off: the surrounding wire has been torn off on one end and a part of the tip of the wire is missing.The missing tip has not been returned with the guidewire.The length estimation of the guidewire concluded in a length of 26-28 mm (33-37 mm according to specification).Therefore, the missing part is considered to have a length of 5-11 mm.The hospital has been informed about the inspection result.The hospital confirmed that no parts remained in the patient.Diagnostics were carried out on the patient during a vascular consultation using imaging.Imaging is understood as x-ray and therefore considered as medical intervention as a preventive measurement to prevent permanent injury.No further harm or clinical consequences have been reported.The case has been changed from a malfunction to a serious injury report.The affected lot number has not been provided by the hospital.A dhr review of relevant lot numbers did not reveal any non conformity or deviation from specification which could have contributed to the event.There is no indication for a systematic root cause as a deficiency of design, production or material considering the very low complaint rate (< 0,01 %, considering all types of picco catheters).Based on the provided information and investigation the root cause is seen in a handling error by the user by withdrawing the guidewire against the needle bevel.According to the problem description "as the guidewire was withdrawn through the cannula, the guidewire was split" and "no problems have been encountered during insertion of the guidewire.But the guidewire could only be withdrawn against resistance.".Withdrawing against the needle, ultimately led to an excessive force, which resulted in the overcoming of the breaking point of the guidewire core and a fracture.The ifu indicates:¿warning: do not withdraw guide wire against needle bevel to avoid possible severing or damage of guide wire.¿ it is not known if the needle has been inserted in a flat angle.The ifu states: "caution: make sure that the introducer cannula is introduced in a flat angle (less than 45°)." according to the customer, the patient had no significant narrowed or calcified arteries, which could have contributed to the event.The ifu indicates: "picco catheters are not intended for any use other than which is indicated.The catheter may not be used in patients where the placement of an indwelling arterial catheter is contraindicated e.G.In the case of arterial prostheses, insufficient perfusion or tissue damages around the puncture site as well as severe peripheral vascular diseases.A picco catheter should only be used if the expected results are reasonable in comparison to the risks." upon the event occurrence the device was involved and catheter insertion procedure on a patient for advanced hemodynamic monitoring was ongoing.The accessory guidewire is supplied in the product picco catheter.The intended user group are trained health care professionals who are used to handling of arterial lines.The seldinger technique is a standard procedure for intended users and is also described in the instructions for use.The issue is monitored on a regularly basis in order to detect early trends.As there is no trend for this kind of issue no additional actions will be taken.With the information stated above the complaint will be closed.
 
Event Description
The manufacture originally reported this complaint as a product problem-malfunction.The following has been reported: as the guidewire was withdrawn through the cannula, the guidewire was split.No harm or clinical consequences for patient or user have been reported.A follow-up was submitted by the manufacturer stating - no problems have been encountered during insertion of the guidewire.But the guidewire could only be withdrawn against resistance.Then on feb 14 2023 new information was received stating: only the guidewire has been returned.Inspection of the returned product has been performed.The microscopic inspection revealed that the guidewire is not complete.The hospital has been informed about the inspection result.The hospital confirmed that no parts remained in the patient.Diagnostics were carried out on the patient during a vascular consultation using imaging.Imaging is understood as x-ray and therefore considered as medical intervention as a preventive measurement to prevent permanent injury.No further harm or clinical consequences have been reported.Manufacturer reference: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PICCO CATHETER
Type of Device
PROBE, THERMODILUTION
Manufacturer (Section D)
PULSION MEDICAL SYSTEMS SE
hans-riedl-strasse 17
feldkirchen 85622
GM  85622
Manufacturer (Section G)
PULSION MEDICAL SYSTEMS SE
hans-riedl-strasse 17
feldkirchen 85622
GM   85622
Manufacturer Contact
christina carlsohn
hans-riedl-strasse 17
feldkirchen 85622
GM   85622
0894599140
MDR Report Key16407710
MDR Text Key309895184
Report Number3013876692-2023-00010
Device Sequence Number1
Product Code KRB
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K171620
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/21/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPV2014L22-A
Device Catalogue Number6885047
Device Lot NumberREQUESTED
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA02/21/2023
Distributor Facility Aware Date02/14/2023
Event Location Hospital
Date Report to Manufacturer02/21/2023
Date Manufacturer Received02/14/2023
Was Device Evaluated by Manufacturer? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
N/A.
Patient Outcome(s) Required Intervention;
-
-