• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: D.O.R.C. DUTCH OPHTHALMIC RESEARCH CENTER INT. EVA TDC VITRECTOMY PACK VGPC INPUT 25G; SURGICAL PACKS AND KITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

D.O.R.C. DUTCH OPHTHALMIC RESEARCH CENTER INT. EVA TDC VITRECTOMY PACK VGPC INPUT 25G; SURGICAL PACKS AND KITS Back to Search Results
Catalog Number 8310.25G12
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/10/2023
Event Type  malfunction  
Manufacturer Narrative
The complaint is under investigation.
 
Event Description
We have been informed that during procedure, the trocar was leaking through the valve.No report that actual harm occurred or surgery was prolonged > 30 minutes.
 
Event Description
We have been informed that during procedure , the trocar was leaking through the valve.No report that actual harm occurred or surgery was prolonged greater than 30 minutes.
 
Manufacturer Narrative
Unfortunately, since the involved trocar was not returned, no physical examination could be performed.Device history record review revealed no deviations and a database search showed that no similar complaints have been reported on this specific lot.However, based on other similar complaints on other lots of this product a detailed investigation of the manufacturing process revealed that when the drying time of the applied glue on the cap of the product is too short for the volume of glue dispensed, valves may in some cases become sticky.Subsequent to the investigation, the drying time was increased as a corrective action to prevent similar issues.Therefore, the most plausible rootcause of this complaint is considered a sticky valve due to insufficient drying time of glue in relation to the volume of glue applied.In 2022 an increase in complaint rate was observed, which triggered a detailed investigation of the manufacturing process.As a result the dry time will be increased, as this was considered the cause of the increase in complaints observed.Per ecf 2022-390, a change was initiated to increase the drying time between glue application and next steps of assembly.The change was implemented with an update of all relevant work instructions.The analysis includes all complaints with failure modes ci-closurevalve-leakage and ci-closurevalve-defect-removal related to comparable trocar systems.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVA TDC VITRECTOMY PACK VGPC INPUT 25G
Type of Device
SURGICAL PACKS AND KITS
Manufacturer (Section D)
D.O.R.C. DUTCH OPHTHALMIC RESEARCH CENTER INT.
scheijdelveweg 2
zuidland, 3214 VN
NL  3214 VN
Manufacturer (Section G)
D.O.R.C. DUTCH OPHTHALMIC RESEARCH CENTER INT.
scheijdelveweg 2
zuidland, 3214 VN
NL   3214 VN
Manufacturer Contact
petra holland
scheijdelveweg 2
zuidland, 3214 -VN
NL   3214 VN
MDR Report Key16416242
MDR Text Key309952521
Report Number1222074-2023-00016
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeAU
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8310.25G12
Device Lot Number2000439340
Is the Reporter a Health Professional? No
Was Device Evaluated by Manufacturer? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-