Model Number 261221
Device Problems Flaked (1246); Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/02/2023
Event Type  malfunction  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported the label of a disposable perforator (id 261221) started to peel and flake off in the patient's brain when drilling.The flakes were removed by using an irrigation, and no patient injury was reported.The device was used for aneurysm clipping.The drill used with the perforator was an electric anspach, the perforator clicked when placed in the drill, and the recommended spring tests were being performed between each burr hole.
 
Manufacturer Narrative
The disposable perforator (id 261221) was returned for evaluation.Device history record (dhr) - there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Failure analysis - the perforator unit was inspected using the unaided eye.Unit #1 was heavily soiled with organic material, and had visible flaking on the eto label.Unit looked as if it had gotten wet, likely from irrigation during surgery, but the cause is unknown.Unit #2 had very light soiling, but showed no other visible anomalies.It can be confirmed the eto label flaked off, however the cause is unknown and cannot be confirmed without further information provided from during the surgery.The information was requested from the hospital but no additional information could be given.Therefore, the complaint condition could be confirmed while the root cause of the failure remains undermined.The root cause is undetermined and was unable to be confirmed in the complaint evaluation.Product was received for analysis and the investigation could not confirm the complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
RAYNHAM
325 paramount drive
325 paramount drive
raynham MA
Manufacturer (Section G)
RAYNHAM
325 paramount drive
raynham MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key16417562
MDR Text Key310335201
Report Number3013886523-2023-00052
Device Sequence Number1
Product Code HBF
UDI-Device Identifier10381780513599
UDI-Public10381780513599
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number261221
Device Catalogue Number261221
Device Lot Number6831847
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received04/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/06/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-