• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. 3.2MMX30MM RNGLC+ ACET DRL BIT; INSTRUMENT, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. 3.2MMX30MM RNGLC+ ACET DRL BIT; INSTRUMENT, HIP Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem Foreign Body In Patient (2687)
Event Date 01/09/2023
Event Type  Injury  
Event Description
It was reported that during surgery with a g7 cup, a drill bit broke and is still in the patient.The surgeon was not able to remove, patient is aware of this, and the hospital made an internal notification.Attempts have been made and no further information has been provided.
 
Manufacturer Narrative
(b)(4).Report source: foreign: netherlands.The customer has indicated that the product will not be returned to zimmer biomet for investigation as it was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
No further information at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Component code: mechanical (g04) ¿ drill.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.2MMX30MM RNGLC+ ACET DRL BIT
Type of Device
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16420111
MDR Text Key310002798
Report Number0001825034-2023-00383
Device Sequence Number1
Product Code PBI
UDI-Device Identifier00880304452442
UDI-Public(01)00880304452442(17)310305(10)257470
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number31-323230
Device Lot Number257470
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/05/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-