• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 2088TC/58
Device Problems Difficult to Insert (1316); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/01/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented for an implant procedure.It was noted that it was difficult to insert the right ventricular lead to the introducer causing the lead to be bent.The lead was not used and replaced.The patient was stable.
 
Manufacturer Narrative
The reported events of difficulty getting the lead through the six french introducer could not be confirmed while lead bent was confirmed.Final analysis found that as received, a complete lead was returned with the helix retracted and clogged blood/tissue.A ring electrode measurement test was performed and the results within specification.The lead test with the introducer could not be performed due to as received lead condition.Visual examination found the helix housing separated from the ring electrode assembly and bent at the distal region consistent with procedural damage.The cause of the reported event of lead bent was isolated to the lead damaged consistent with procedural damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16433885
MDR Text Key310158308
Report Number2017865-2023-10785
Device Sequence Number1
Product Code NVN
UDI-Device Identifier05414734502894
UDI-Public05414734502894
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2088TC/58
Device Catalogue Number2088TC-58
Device Lot NumberA000135014
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/06/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age51 YR
Patient SexMale
-
-