DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problems
Activation, Positioning or Separation Problem (2906); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 02/14/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that, the cs300 intra-aortic balloon pump (iabp) unit loses operational capabilities.
|
|
Manufacturer Narrative
|
Updated fields - e1(site country), h6 (medical device ¿ problem code , type of investigation, investigation findings, investigation conclusions).Getinge field service engineer (fse) tested iabp with mini sim and tested iabp invasive and static pressure and passed to specifications.Fse could not duplicate failure observed by end user.Unit passed all functional and safety tests per factory specifications, returned to customer and cleared for clinical use.
|
|
Event Description
|
It was reported that, the cs300 intra-aortic balloon pump (iabp) unit loses operational capabilities.Information about patient involvement and harm/injury is unknown.
|
|