OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problems Display Difficult to Read (1181); Display or Visual Feedback Problem (1184)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Date of event and device identification (udi number) is unknown.No information has been provided to date.Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the display was missing a segment.There has been no report of patient involvement, no observable clinical symptoms, or a change in symptoms identified in the patient.
 
Manufacturer Narrative
Other text: h6: event methods, evaluation, and conclusion codes: updated.One warmer device was received for investigation.During visual inspection the device was observed to have a damaged tank cover and enclosure.Corroded drain fitting and faded line cord.The circuit board and power switch were outdated.The reported issue was confirmed during functional testing when the warmer display was missing lines of pixels.However the investigation could not identify a root cause for the observed condition of the display.As no manufacturing root cause could be identified, no review of manufacturing device history records was conducted.A review of device confirmed this device has not been in for service previously.Due to the age of the device, it has been deemed beyond economical repair and will be decommissioned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key16439745
MDR Text Key310254672
Report Number3012307300-2023-01669
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? No
Date Returned to Manufacturer02/15/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/28/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-