Model Number D97130F5 |
Device Problem
Failure to Capture (1081)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/02/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Without the return of the product, it is not possible to determine if damages or defects exist on the product, nor can any manufacturing nonconformance, failure mode, root cause, or potential contributing factors be identified.The lot number was not provided; therefore, a review of the manufacturing records could not be completed.An engineering evaluation has been initiated to assess for any manufacturing-related processes which could be correlated to the complaint.Complaint histories for all reported events are reviewed against trending control limits and any excursions above the control limits are assessed and documented as part of this monthly review.
|
|
Event Description
|
It was reported that a swan ganz model d97130f5, lot unknown, was inserted into right ventricle from the right femoral vein.It was plugged into the external temporary pacemaker and the temporary pacemaker was turned on.When the temporary pacemaker was turned on, and effort to pace the right ventricle, there was no stimulus observed on the ekg.Multiple attempts were made to reposition the catheter in the right ventricle.Regardless of where the catheter was positioned in the rv, the team was never able to obtain capture.The temp pacing catheter was removed, and a new edwards pacing catheter was inserted and tested successfully.The second temporary pacing catheter functioned as expected, without issue.There was no patient injury.
|
|