BOSTON SCIENTIFIC CORPORATION INTERLOCK-35; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
|
Back to Search Results |
|
Model Number 83787 |
Device Problems
Premature Activation (1484); Failure to Advance (2524); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/27/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device evaluated by mfr.: visual inspection was performed.Only the main coil returned for the product analysis.The main coil returned kinked, stretched and detached in the interlocking arm.Functional inspection, since only the main coil returned for product analysis the functional inspection could not to be performance.
|
|
Event Description
|
Reportable based on device analysis completed on 09feb2023.It was reported that coil advancing issue occurred.The target lesion was located in the left renal artery.A 10mm x 40cm interlock cube coil was selected for use.During the procedure, the coil was found unloaded and stuck in the catheter.The device was unable to be pushed after several attempts.The procedure was completed by another of the same device.No patient complications were reported and the patient status was stable.However, returned device analysis revealed a detached interlocking arm.
|
|