• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTERLOCK-35; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTERLOCK-35; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number 83787
Device Problems Premature Activation (1484); Failure to Advance (2524); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/27/2022
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr.: visual inspection was performed.Only the main coil returned for the product analysis.The main coil returned kinked, stretched and detached in the interlocking arm.Functional inspection, since only the main coil returned for product analysis the functional inspection could not to be performance.
 
Event Description
Reportable based on device analysis completed on 09feb2023.It was reported that coil advancing issue occurred.The target lesion was located in the left renal artery.A 10mm x 40cm interlock cube coil was selected for use.During the procedure, the coil was found unloaded and stuck in the catheter.The device was unable to be pushed after several attempts.The procedure was completed by another of the same device.No patient complications were reported and the patient status was stable.However, returned device analysis revealed a detached interlocking arm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLOCK-35
Type of Device
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key16452023
MDR Text Key310331295
Report Number2124215-2023-08780
Device Sequence Number1
Product Code KRD
UDI-Device Identifier08714729795513
UDI-Public08714729795513
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K133208
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 02/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number83787
Device Catalogue Number83787
Device Lot Number0028802517
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexFemale
Patient Weight65 KG
-
-