Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional device product codes: gfa, gff, hwe complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.A device history record (dhr) review was conducted.Part: 310.250 lot: 5l22644 manufacturing site: (b)(4).Release to warehouse date: 26 july 2019.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.H6: photo investigation: the photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned photo.Visual analysis of the photo revealed that the edges of the flutes of drill bit ø2.5 l110/85 2flute f/quick co appear to be softened and dull.The observed condition was consistent as an end of life indicator for the device.No other issues were identified.After a visual inspection, it was determined that the reusable instrument device was dullness from repeated use and servicing; therefore, further investigation for the reported complaint device is not required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed dullness condition of drill bit ø2.5 l110/85 2flute f/quick co would have contributed to the complained issue.There is no indication that a design or manufacturing issue has caused the complaint condition.It was determined that the reusable instrument was worn from repeated use and servicing.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|