There was an allegation of questionable elecsys ferritin results for 2 patient samples on a cobas e 411 immunoassay analyzer compared to a siemens centaur analyzer.For sample 1, the e411 ferritin result was 373.6 ng/ml.The repeat result on the centaur analyzer was 226 ng/ml.For sample 2, the e411 ferritin result was 760 ng/ml.The repeat result on the centaur analyzer was 506 ng/ml.No questionable results were reported outside of the laboratory.The e411 analyzer serial number was requested but not provided.
|
The e411 serial number was (b)(6).The qc recovery data provided was acceptable.Based on the available data, a reagent issue could be excluded.The investigation determined that the result differences generated between the different methods are consistent with methodological differences.This relates to the overall setups of the assays, the antibodies used, differences in reference materials/methods, and the standardized methodology used.The investigation did not identify a product problem.The root cause of the event could not be determined.
|