• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number LIFEVEST WCD 4000 SYSTEM
Device Problem Communication or Transmission Problem (2896)
Patient Problem Insufficient Information (4580)
Event Date 01/24/2023
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of monitor has been completed.The reported problem (does not communicate with belt) was due to the monitor's electrode belt connector being broken free from the monitor enclosure, damaging wires within the connector.The root cause for the damaged connector was excessive force.There was no adverse event that resulted from the damaged monitor.
 
Event Description
A us distributor returned a monitor and reported being unable to communicate with belt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA
Manufacturer Contact
zoll mfg corporation
121 gamma drive
pittsburgh, PA 15238
MDR Report Key16458920
MDR Text Key310410659
Report Number3008642652-2023-01956
Device Sequence Number1
Product Code MVK
Combination Product (y/n)N
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 03/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberLIFEVEST WCD 4000 SYSTEM
Date Manufacturer Received02/23/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-