In the morning, patient began coughing quite a bit, which had been occurring often over the past several days.During the previous shift that i had cared for the patient, i had to change the proximal flow sensor on the hamilton g5 vent 3 times due to excessive secretions in the flow sensor.The vent circuit with the sensor was in the appropriate position per manufacturer's recommendations to prevent secretion interference.The sensor was producing alerts for high and low tidal volumes, high and low minute volumes, check flow sensor and turn flow sensor alarms.Approximately 10 minutes later, patient started to desaturate to 70's with coughing and his respirations climbed to 50's despite 100% fio2 on ventilator.Writer removed patient from vent and manually ventilated to 100% spo2.Patient was returned to vent and immediately began to drop oxygen saturations again, while also receiving sedation.Once again, patient was manually bagged to 100%.Upon returning to ventilator, and after patient was also chemically paralyzed, the vent wouldn't cycle and deliver a breath.Writer immediately removed patient from ventilator & resumed bagging and called for a new ventilator.
|