STERIS MEXICO, S. DE R.L. DE C.V. V-PRO MAX STERILIZER Back to Search Results
Device Problem Leak/Splash (1354)
Patient Problem Skin Burning Sensation (4540)
Event Date 02/01/2023
Event Type  malfunction  
Manufacturer Narrative
A steris service technician arrived onsite following the reported event to inspect the v-pro max sterilizer and found that the injection cylinder was leaking peroxide which then dripped onto the power switch.The technician replaced the injection cylinder, tested the unit, found it to be operating properly, and returned it to service.A three-year complaint review indicates this to be an isolated event.No additional issues have been reported.
 
Event Description
The user facility reported that an employee experienced skin irritation on their hand after coming in contact with a wet substance while powering down their v-pro max sterilizer.The employee washed their hands and continued working.No medical treatment was sought or administered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-PRO MAX STERILIZER
Type of Device
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key16460369
MDR Text Key310431989
Report Number3005899764-2023-00010
Device Sequence Number1
Product Code MLR
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/01/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/13/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse