Model Number 394.46 |
Device Problems
Device-Device Incompatibility (2919); Material Twisted/Bent (2981)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/10/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint #(b)(4) depuy synthese is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.Reporter is a j&j sales representative.The subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that on (b)(6) 2023, the holding sleeve 105mm has a cross threaded nut that is stuck and sddrive screwdriver shaft t15 250 long gets stuck in quick connect.There was no operation involved.There was no patient involvement.It was unknown when was the event occurred.There was no patient consequences.This complaint involves two (2) devices.This report is for one (1) holding sleeve 105mm this is report 2 of 2 for complaint (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A manufacturing record evaluation (mre), was not possible because the required lot code was not provided.Service and repair evaluation: the customer reported that for the device 05.000.008, hand piece for battery powered driver, doesn¿t have power.The repair technician reported that the wing screw is cross threaded into the holding sleeve body, unable to remove wing nut, therefore the unit cannot be repaired.Damaged component is the reason for repair.The item is not repairable per the inspection sheet.The cause of the issue is unknown.The item will be forwarded to customer quality.The evaluation was confirmed.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|