The reported events of no low voltage (lv) output and failure to interrogate were confirmed.The device was received with no telemetry communication and no output.Visual inspection of the header attachment area detected an anomaly between the pre-molded header and titanium case.The device was cut open to enable further testing and the battery was found at normal voltage levels.Feedthrough leak test was performed, indicating a device hermeticity breach.This is consistent with feedthrough damage as a result of fluid intrusion between the header and case, and subsequent fluid ingress to internal electronics.Hybrid circuitry was tested, and the result indicated normal current drain.A manufacturing process anomaly may have occurred, which resulted in the header bonding anomaly.As a result of this finding, abbott is performing further investigation.The device is included in the zenex, assurity, endurity laser adhesion preparation advisory issued by abbott on 20 july 2022 for a subset of devices distributed and implanted outside of the united states.
|