OAKDALE LEVEL 1 HOTLINE BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Catalog Number CON-HL-90
Device Problem Defective Device (2588)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Udi number not available.Date of event, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.
 
Event Description
It was reported that the system failed.Date of event, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.
 
Manufacturer Narrative
Other text: h3.Device evaluated by manufacturer and h6.Health impact, event problem and evaluation codes: updated.One (1) device was received.Visual inspected noted a cracked tank cover.Outdated printed circuit board (pcb) and power switch.Filled tank with water, attached temp check, plugged in line cord, and turned on the power switch.The system failed in that it wouldn't even power on confirming the customer complaint.A root cause was the connection between the power switch and printed circuit board (pcb) was damaged.A manufacturing device history record (dhr) review was not performed because was no indication of a manufacturing defect during the investigation.Service history review identified this device has not been in for service previously.No action taken due to the age and condition of the device.It is deemed beyond economical repair and will be scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE BLOOD AND FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key16468706
MDR Text Key310593122
Report Number3012307300-2023-01910
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? No
Date Returned to Manufacturer02/22/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/20/2001
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-