Lot Number 0030456962
Device Problems Difficult to Flush (1251); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/03/2023
Event Type  malfunction  
Event Description
During a pvi cryoablation procedure a polarsheath was selected for use.Due to improper packing, the flexible pipe has been bent/squeezed too much, so the liquid (saline) was not able to pass through.An exchange of the polarsheath resolved the issue.Te procedure was completed without any patient complications.
 
Manufacturer Narrative
The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARSHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
302 parkway
global park, la aurora
heredia
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key16469070
MDR Text Key310530469
Report Number2124215-2023-09501
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodePL
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/11/2023
Device Lot Number0030456962
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/31/2022
Is the Device Single Use? Yes
Type of Device Usage Unknown