• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. COMP AUG MINI BSPLT W TPR SM; PROSTHESIS, SHOULDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. COMP AUG MINI BSPLT W TPR SM; PROSTHESIS, SHOULDER Back to Search Results
Model Number N/A
Device Problems Material Erosion (1214); Detachment of Device or Device Component (2907)
Patient Problems Erosion (1750); Failure of Implant (1924); Pain (1994); Joint Dislocation (2374); Osteopenia/ Osteoporosis (2651); Metal Related Pathology (4530)
Event Date 02/05/2023
Event Type  Injury  
Event Description
It was reported that the patient underwent a right reverse total shoulder arthroplasty.Approximately 1 month post implantation, the patient was revised due to pain, scapular notching, and dissociation of the glenosphere from the baseplate as noted on x-ray.During the revision, severe metallosis was noted throughout the joint due to severe wear of the poly and baseplate as well as the dislocation and disassociation.Significant bone loss and erosion were also noted.The stem was left intact, and the glenoid was converted to a hemiarthroplasty with a new head and taper placed.
 
Manufacturer Narrative
(b)(4).Product code - (b)(4).110031402 - humeral tray - 64437435.110031424 - humeral bearing - 64833589.115310 - glenosphere - 290790.180551 - screw - 294290.180553 - screw - 380950.180553 - screw - 006800.180550 - screw - 758340.115395 - central screw - 071640.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2023-00421.
 
Manufacturer Narrative
No further event information available at the time of this report.H10: this follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records and radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: office visit: presents with pain, specifically in the anterior region of the shoulder.The patient reports sleeping in her recliner when she felt the pain approximately 2 months prior.X-rays show dissociation of the glenosphere from the baseplate with no apparent loosening of the humeral component or baseplate.Scapular notching is present with multiple loose bodies around the joint.Plan for revision.Revision, hemiarthroplasty: revision due to dislocation and disassociation of the glenoid components.Severe metallosis noted throughout the joint.Severe wear of the inferior aspect of the poly as well as some wear of the metal of the humeral tray.Humeral stem found stable.Erosion of the cortical bone at the lesser tuberosity with stem visible through a cortical window.Severe wear of the baseplate noted with the central screw hole also malformed.¿there had been melting of the posterior aspect of the baseplate and then it had reached annealed with metal covering the posterior locking screw in the baseplate.This made that screw inaccessible.¿ ¿between the patient¿s pre-existing glenoid bone deformity as well as the bone loss with removal of the failed baseplate, there was not enough remaining native glenoid bone to perform a reverse shoulder arthroplasty even with a large augment.¿ hemi components placed without complication.The patient to remain non-weight bearing with no shoulder motion for 4 weeks then proceed with rom exercises.X-rays: reverse total shoulder arthroplasty of the right shoulder with implant disassembly of the glenosphere, now located along the lateral aspect of the humeral component.Bony remodeling of the distal clavicle and ac joint, acromion with extensive curvilinear calcification of the joint capsular which presumably relates to patient's known history of metallosis.Osteopenia is present.Central screw home affirmation cannot be confirmed on x-ray.Taper adapter not clearly seen.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP AUG MINI BSPLT W TPR SM
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16474082
MDR Text Key310588255
Report Number0001825034-2023-00422
Device Sequence Number1
Product Code PHX
UDI-Device Identifier00880304993235
UDI-Public(01)00880304993235(17)240424(10)64394791
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172502
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/02/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/24/2024
Device Model NumberN/A
Device Catalogue Number110032410
Device Lot Number64394791
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/24/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE.
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexFemale
Patient Weight67 KG
-
-