• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM; MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM; MITRAL VALVE REPAIR DEVICES Back to Search Results
Catalog Number CDS0705-XTW
Device Problems Break (1069); Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/09/2023
Event Type  malfunction  
Event Description
This will be filed to report a cracked clip introducer.It was reported that during device prep of a mitraclip procedure, while the clip delivery system (cds) was being inserted into the clip introducer, the clip introducer was entered between the clip and the shaft.Subsequently the clip introducer cracked.The cds was exchanged and the procedure was completed without further complication.There was no clinically significant delay in the procedure and no patient involvement.No additional information was provided.
 
Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Manufacturer Narrative
All available information was investigated, and the reported cracked clip introducer was confirmed.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents reported from this lot.Based on available information, the reported cracked clip introducer was due to procedural circumstances.There is no indication of a product quality issue with respect to manufacture, design or labeling.H6: code 1069 removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key16474655
MDR Text Key310822817
Report Number2135147-2023-00980
Device Sequence Number1
Product Code NKM
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/16/2023
Device Catalogue NumberCDS0705-XTW
Device Lot Number21117R1091
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/17/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-