MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD AUTOFILL CHAMBER; BTT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD AUTOFILL CHAMBER; BTT Back to Search Results
Model Number 900PT290E
Device Problem Overfill (2404)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
A distributor in india reported via a fisher and paykel healthcare (f&p) field representative that a 900pt290e autofill chamber with adapter was found to have overfilled above the maximum water level line during use.There were no reported patient consequences.
 
Manufacturer Narrative
(b)(4).Fisher & paykel healthcare (f&p) is currently in the process of completing our investigation.We will provide a follow up report upon completion of our investigation.Product background: the 900pt290e autofill chamber with adapter (900pt290e) is a component designed for use with the airvo 2 humidification series to provide nasal high flow (nhf) therapy.Air and oxygen are blown from the airvo 2 into the humidification chamber within which water is heated by a heater plate.The gases become humidified in the chamber and then pass into a heated breathing tube, then to the patient interface.Nhf therapy is intended for use with spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases nhf therapy should not be used for life support purposes, and appropriate patient monitoring must be used at all times.
 
Manufacturer Narrative
(b)(4).Product background: the 900pt290e autofill chamber with adapter (900pt290e) is a component designed for use with the airvo 2 humidification series to provide nasal high flow (nhf) therapy.Air and oxygen are blown from the airvo 2 into the humidification chamber within which water is heated by a heater plate.The gases become humidified in the chamber and then pass into a heated breathing tube, then to the patient interface.Nhf therapy is intended for use with spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases nhf therapy should not be used for life support purposes, and appropriate patient monitoring must be used at all times.Method: the complaint 900pt290e was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is based on the information and videography provided by the customer, previous investigations of similar complaints, and our knowledge of the product.Method: visual inspection of the provided videography revealed that the chamber had overfilled above the maximum water level line.Conclusion: without the return of the complaint device, we are unable to confirm the cause of the reported event.Every 900pt290e has their float functions and valves tested during the manufacturing process.Any chamber that fails this inspection is rejected.The subject chamber would have met the required specification at the time of production.The user instructions which accompany the 900pt290e as part of the 900pt561 heated breathing tube and chamber kit show in pictorial format the correct water level for the 900pt290e.The 900pt561 heated breathing tube and chamber kit also state the following: "for single patient use only.Reuse may result in transmission of infectious substances.Attempting to reprocess will result in degradation of materials and render the product defective." "do not use the chamber if it has been dropped or been allowed to run dry as this could lead to over-filling." "do not use the chamber if the water level rises above the maximum water level line as this may lead to water entering the patient's airway." "avoid contact with chemicals, cleaning agents, or hand sanitizers.".
 
Event Description
A distributor in india reported via a fisher and paykel healthcare (f&p) field representative that a 900pt290e autofill chamber with adapter was found to have overfilled above the maximum water level line during use.There were no reported patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOFILL CHAMBER
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite 300
irvine, CA 92618
9494534002
MDR Report Key16475595
MDR Text Key310698360
Report Number9611451-2023-00151
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K131895
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number900PT290E
Device Catalogue Number900PT290E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
F&P 900PT561 HEATED BREATHING TUBE AND CHAMBER KIT; F&P 900PT561 HEATED BREATHING TUBE AND CHAMBER KIT; F&P OPT944 OPTIFLOW + ADULT NASAL CANNULA; F&P OPT944 OPTIFLOW + ADULT NASAL CANNULA; F&P PT101 AIRVO 2 HUMIDIFIER; F&P PT101 AIRVO 2 HUMIDIFIER
-
-