• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH MATRIX 1.1MM DRILL BIT J-LATCH/8MM STOP/44.5MM; DRILL, BONE, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH MATRIX 1.1MM DRILL BIT J-LATCH/8MM STOP/44.5MM; DRILL, BONE, POWERED Back to Search Results
Catalog Number 03.503.248
Device Problems Break (1069); Entrapment of Device (1212)
Patient Problem Foreign Body In Patient (2687)
Event Date 02/10/2023
Event Type  Injury  
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Procodes: erl and hbe.Device available for evaluation: complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.The initial reporter is a j&j sales representative.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in macau as follows: it was reported on (b)(6) 2023, that drill bit got broken while drilling the third drill hole for the fixation screw for fix the cutting guide.Hospital had took x-ray and found the drill bit tip stuck in patients¿ zygomatic bone.Finally the tips cannot be take out, as stuck in patient's bone.The surgery was delayed for 120- 150 minutes due to the reported event.The procedure was successfully completed.Fragments were generated.Were they removed easily without additional intervention? no.This report is for one (1) drill bit ø1.1 w/stop l44.5/8.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: part: 03.503.248.Synthes lot: u352768.Supplier lot: u352768.Release to warehouse date: 22 december 2019.Supplier: (b)(4) no nonconformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the drill tip of the drill bit ø1.1 w/stop l44.5/8 was found broken, the broken fragments were not returned.The device, component or fragment remains in patient issues could be not confirmed since x-ray evidence was not provided.A dimensional inspection was not performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.The overall complaint was confirmed as the observed condition of the drill bit ø1.1 w/stop l44.5/8 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Drawing/specifications reviewed? current and manufactured revisions were reviewed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MATRIX 1.1MM DRILL BIT J-LATCH/8MM STOP/44.5MM
Type of Device
DRILL, BONE, POWERED
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16477715
MDR Text Key310606832
Report Number8030965-2023-02599
Device Sequence Number1
Product Code DZI
UDI-Device Identifier07611819901513
UDI-Public(01)07611819901513
Combination Product (y/n)N
Reporter Country CodeMC
PMA/PMN Number
K043310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/03/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.503.248
Device Lot NumberU352768
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/22/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNK - SCREWS: TRAUMA.
Patient Outcome(s) Required Intervention;
-
-