Model Number 90620US
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The duo headlight 2 bay system (90620us) was returned for evaluation: failure analysis - the returned duo headlight (90620us) was received is in used condition.Evaluation found that the unit's power cord was losing connection, leading to the fan malfunction and also leading to the light flickering.The power cord will be replaced to address this issue.Root cause - evaluation identified that the reported issue was most likely due to routine use and wear.The reported complaint was confirmed.No manufacturing, workmanship, or material deficiency has been identified.
 
Event Description
A facility reported that the light of duo headlight 2 bay system (90620us) was flickering on and off.Additionally, the led lamp seemed to be getting excessively hot.No patient injury, death or surgical delay was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO HEADLIGHT 2 BAY SYSTEM - US
Type of Device
PFM05
Manufacturer (Section D)
INTEGRA LIFESCIENCES (IRELAND) LTD.
sragh tullamore, co. offaly
EI 
Manufacturer (Section G)
INTEGRA LIFESCIENCES (IRELAND) LTD.
sragh tullamore, co. offaly
EI  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key16477784
MDR Text Key310625865
Report Number3006697299-2023-00021
Device Sequence Number1
Product Code FSR
UDI-Device Identifier10381780490746
UDI-Public10381780490746
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/03/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number90620US
Device Catalogue Number90620US
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2023
Date Manufacturer Received02/08/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-