We received an allegation of questionable results for 1 patient's sample tested with the trigl assay on cobas pro c 503 analytical unit serial number (b)(4).On (b)(6) 2023 at 8:27 a.M.The customer ran 1 patient's sample and initially resulted in trigl value of 266 mg/dl.The result was reported to the physician but it did not correlate with the patient's clinical picture so he asked to retest the sample again.On the same day at 10:45 a.M.The patient's sample was repeated from the same primary tube and on the same analyzer and resulted in trigl value of 143 mg/dl.Later in the day, the patient's sample was repeated but this time on another c503 module, and resulted in trigl value of 144 mg/dl.
|