LIVANOVA DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 10-85-00 |
Device Problem
Pumping Stopped (1503)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/06/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova deutschland manufactures the s5 double head pump.The incident occurred in tottori, japan.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland received a report that a s5 double head pump gave twice an error message related to shaft angle when the rotation speed pump was stopped.The issue occurred during procedure.There was no patient injury.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H10: a livanova field service representative was dispatched to the facility to investigate the device and could not reproduce the reported issue.However, in the serial read-out of the pump (real time device parameters and setting recording file) shaft angle encoder error, that occurred twice, was stored.In order to solve the issue, shaft angle encoder was replaced.Subsequent functional verification testing was completed without further issues and the unit was returned to service.Complaints database analysis revealed that no similar event on this device occurred since its installation in 2015.It cannot be ruled that an electrical/electronic deviation of the replaced component was the root cause of the reported issue.
|
|