The e411 serial number was (b)(6).The calibration trace's signal counts were within specifications.The qc results provided do not reflect the relevant time.However, according to the technical support, the calibration and qc were both acceptable on the day of testing.Limited pre-analytical information was provided.The alarm trace data was not provided, however, it was reported that the alarm trace did not indicate any issues.Based on the data provided, a general reagent issue can be excluded.The investigation did not identify a product problem.The cause of the event could not be determined.
|