|
Model Number 8668 |
Device Problems
Poor Quality Image (1408); Retraction Problem (1536); Device Displays Incorrect Message (2591)
|
Patient Problem
Non specific EKG/ECG Changes (1817)
|
Event Date 02/16/2023 |
Event Type
Injury
|
Event Description
|
It was reported that st elevation occurred.The target lesion was located in a mildly tortuous and moderately calcified no stenosis left anterior descending artery (lad), no stenosis circumflex (cx), and 70% stenosed left main (lm) before the rotablation.After left main rotablation completed, the opticross hd catheter used to check the result.After washing the catheter several times, the catheter was inserted into the lad, the pullback started but the image was incomprehensible.Then, the pullback stopped, and error message appeared "the image cannot be ealuated".The circumflex (cx) was the same and st elevation was noted.The physician removed the device and the procedure canceled.The patient was stable post procedure.
|
|
Manufacturer Narrative
|
Device evaluated by manufacturer: visual inspection of the device revealed that the telescope and the sheath were observed kinked.No damages or failures were observed on the ccp (catheter code pcba) board assembly.Microscopic inspection noted pitting and degradation of the transducer housing consistent with saline damage.The mdu and ilab system were used to perform a functional inspection on the device.The catheter was properly recognized by the imaging system when plugged into the mdu and no connection issues or errors were detected during its testing.It was observed that the catheter flushed normally when the imaging core was fully retracted and fully advanced.Imaging core impedance test shows a complete circuit curve.Transducer level impedance test shows a good rh peak of 40 ohms.During image characterization testing, a good square image appeared in the ilab system.Auto pullback was not performed with the sled due to the kink in the telescope.The telescope assembly was able to properly pull back, advance, and retract; however, some resistance was found due to the kink in the telescope.Inspection of the remainder of the device, revealed no other damage or irregularities.
|
|
Event Description
|
It was reported that st elevation occurred.The target lesion was located in a mildly tortuous and moderately calcified no stenosis left anterior descending artery (lad), no stenosis circumflex (cx), and 70% stenosed left main (lm) before the rotablation.After left main rotablation completed, the opticross hd catheter used to check the result.After washing the catheter several times, the catheter was inserted into the lad, the pullback started but the image was incomprehensible.Then, the pullback stopped, and error message appeared "the image cannot be evaluated".The circumflex (cx) was the same and st elevation was noted.The physician removed the device and the procedure canceled.The patient was stable post procedure.
|
|
Search Alerts/Recalls
|
|
|